Home > Products > Imaging systems > Multifunctional fluorescence & chemiluminescence imaging analysis system (BG-gdsAUTO 730)

Multifunctional fluorescence & chemiluminescence imaging analysis system (BG-gdsAUTO 730)

BG-gdsAUTO 730
High sensitivity

BG-gdsAUTO730 multifunctional fluorescence chemiluminescence imaging system is suitable for various applications such as multi-color fluorescence imaging analysis, chemiluminescence imaging analysis, nuclear acid gel imaging analysis and protein gel imaging analysis.

  • Introduction
  • Features
  • Parameters
  • Support
    • BG-gdsAUTO730 multifunctional fluorescence chemiluminescence imaging system is suitable for various applications such as multi-color fluorescence imaging analysis, chemiluminescence imaging analysis, nuclear acid gel imaging analysis and protein gel imaging analysis. The first class high-resolution low-illumination digital refrigeration camera ensure image quality; The large-aperture, high-throughput, fixed-focus lenses greatly improve detection sensitivity to detect weak signals; The deep-cooled CCD with low background noise improve image quality even in case of long exposure time. 6-bit electric filter wheel to meet a variety of experimental needs.

    • 1.       **Integrated design of multi-color fluorescence such as Cy2, Cy3, Cy5 imaging, chemiluminescence imaging and ordinary gel imaging analysis, truly multi-purpose machine, saving experimental space and easy to use;

      2.       *No dark room required: Instead of dark room exposure, no dark room is wasted, no tedious fumbling exposure steps, no contact with toxic chemicals;

      3.       *High sensitivity: scientific research CCD, high quantum efficiency, high detection sensitivity, cooling CCD imaging system reduce background noise;

      4.       *Large aperture lens: F0.8, collect more light to improve detection sensitivity;

      5.       *Automaitic exposure and estimating the exposure time in real time;

      6.       *Multiple imaging modes: single/ series/ manual;

      7.       *Wide dynamic range: detect weak bands alongside strong bands.


      Application

      Fluorescence detection such as Cy2, Cy3, Cy5;

      Chemiluminescence imaging of Western blot membrane with various luminous substrates such as Western Lightning, ECL, ECL plus, CDP Star, SuperSignal, CSPD, LumiGlo;

      DNA / RNA gel imaging;

      Protein gels imaging.


    • Camera: high resolution low illumination digital refrigeration camera

      Resolution: 2750×2200(6.05million)

      Pixel size:4.54μm × 4.54μm

      A/D: 16bit

      Binning:1×1,2×2, 4×4,6×6, 8×8,12×12, 24×24

      CCD photo sensitive chip:SONY ICX 695

      CCD dark current: < 0.0005e-/pixel/sec @-40

      Quantum efficiency: >75%

      CCD cooling method: Semiconductor refrigeration

      CCD cooling temperature: -55°C

      Lens: F/0.8 fixed focus lens

      Zoom: compatible with sample thickness of 0.1-10mm

      Dynamic range: >4.5 orders of magnitude

      Six-position motorized filter wheel: equipped with 535nm, 590nm, 605nm, 700nm , 840nm(optional) filters can also be customized

      Epifluorescence on both sides: red 625nm, green 525nm, blue 470nm,740nm(optional)

      Transmission light source: 302nm

      Reflected light source: UV254nm, UV365nm (optional)

      White light on both sides: 12v LED light source on both sides

      Equipped with professional UV to white light board

      UV transmission area: 20 × 20cm

      UV to white light plate transmission area: 26 × 21cm

      UV reflection area: 15cm × 5cm

      Chemiluminescence imaging area: 15 × 18cm

      Dimensions(L × W × H): 405×440×680mm

      Software: Baygene professional camera software, switch between Chinese and English, easy to operate, free upgrade for life